Wuxi ASK Pneumatic Machinery Co.,Ltd , https://www.askpneumtic.com
Main Technical Specifications
Power Supply
Voltage
Single Phase AC 180~240V
Frequency
50Â HzÂ
Current (Max., instantaneous)
<20A
Power capacity
greater than 5.5kvA
Internal resistance of power supply wire ≤1Ω
Maximum rated capacity
Current(mA)
Voltage (kvp)
Exposure time(s)
30
90
0.08~6.3
40
90
0.08~4.0
50
90
0.08~1.0
80
0.08~2.5
  Timer
0.08s to 6.3s, in 20 shifts, electronic type
  Max distance from tube focus to the groundÂ
≥1820 mm
  Min distance from tube focus to the ground
≤550 mm
  Remote control exposure distance
≥ 6m
  Max. exposure fieldÂ
350×350 mm (when distance between X-ray tube focus and film is 650 mm)
  Max. exposure fieldÂ
450×450 mm / when distance 1000mm
  Specification of x ray tube
XD1-3/100, stationary anode, focus 2.3 x 2.3
Medical 50mA Mobile X-ray Machine (WHY50)
Model NO.: WHY50
Alarm Form: Display Alarm
Packing Dimension: 135cm*75cm*150cm 1.52m³
Voltage: Single Phase AC 180~240V
Frequency: 50 Hz
Current (Max., Instantaneous): <20A
Trademark: Poweam
Transport Package: 1PCS/CTN
Specification: ISO CE
Origin: China
HS Code: 90189090
1. Used for X-ray radiography in wards, especially in orthopaedic wards.
2. Single focus point, high voltage silicon stack full-wave rectification.
3. kV, mA and s capacity interlocking control system, overload indicating system.
4. Wired manual exposure & remote control exposure
5. Sound and light indication during the course of exposure. Â